About The Device

Technical Specs

ParameterValue

Measured Quantity

H*(10) — Ambient dose equivalent rate

Units

Sv, rem, CPS / CPM

Detector

Plastic scintillator 14×14×14 mm with SiPM (temperature compensated)

Alarming

Vibration, LEDs, audible sound

Alarm thresholds

6 (2 each for CPS, dose rate, dose)

Type of radiation

Photons (gamma, X-ray)

Energy range (dose measurement)

100 keV — 3.0 MeV

Energy range (indication)*

30 keV — 3.0 MeV

Count rate range (Cs-137)

0 — 144 MCPS

Dose equivalent value

0.01 µSv/h — 9000 mSv/h

Dose rate accuracy

±20%

Accumulated dose

0—100 Sv, accuracy ±20%

Sensitivity Cs-137 (662 KeV)

16 CPS/(µSv/h)

Sensitivity Co-60 (1250 KeV)

8 CPS/(µSv/h)

Display update period

0.5 s

Memory

≈1000 hours (non-volatile)

Battery life

Up to 200 hours

Battery

Li-Pol 3.7 V, 1500 mAh

Charging: voltage / max current / time

5 V / 750 mA / ≈4 h

Max USB consumption

800 mA

Protection

IP64 (dust and splashing water)

Operating temperature

−20…+45 °C

Max humidity

85%

Wireless

BLE 5.2

Wired

USB 2.0+, Type-C

Display

Monochrome LCD 128×48, 34×13 mm, FSTN, Transflective, Positive

DImensions (W×H×D)

127×34×21 mm

Weight

76 g

* In the 30–100 keV range, the device detects radiation, but dose rate readings may deviate from actual values due to the detector’s energy response characteristics. In other words, the device will detect the presence of radiation but will measure it with greater uncertainty than in the primary range of 100 keV — 3.0 MeV.

System Requirements

Computer: Windows 8/10/11, 2 GB RAM or more, 1 GB free disk space, USB port, 1024×768 display or better.

Android: version 6.0 or higher.

iOS/macOS: iOS 17.0+ / macOS 14.0+ (Mac with Apple Silicon).

Aktualisiert: 8. Apr. 2026