Alarm system

Setting alarm thresholds

Two alarm thresholds can be set for each of the three quantities — count rate, dose rate, accumulated dose. The procedure is the same for all three; below it is shown using count rate as an example. The setting is made from the “Settings” submenu (for access, see chapter “Settings”).

Grid Item
  • Navigate to the quantity — for example, [Count Rate].
  • Enter the submenu.
Grid Item
Grid Item
  • Enter [Alarm 1].
  • Select a digit, change its value.
  • Set all digits.
Grid Item
Grid Item
  • Select [Enter] or [Esc].
  • Long-press to confirm.
Grid Item
Grid Item
  • Navigate to [Alarm 2]. Same procedure.
Grid Item
Grid Item
  • To go back: [< Back] or [<< Menu quit].
Grid Item

Other quantities — same procedure:

Grid Item
  • [Dose Rate] → [Alarm 1] / [Alarm 2] — dose rate alarm thresholds.
Grid Item
Grid Item
  • [Dose] → [Alarm 1] / [Alarm 2] — accumulated dose alarm thresholds.
Grid Item

Updated: Sep 22, 2026